Specifications
100 Calibration and Preventive Maintenance  Chapter 6
3. Select the OK button in the prompt box. Note, if you select the 
CANCEL button, this portion of the Agilent E6171A calibration is 
not performed.
4. When prompted by the ‘autoadj.tpa’ testplan , remove the test 
fixture ribbon cable and reconnect the ribbon cable from the E6172A 
Matrix Module.
5. Select the OK button in the prompt box.
6. The ‘autoadj.tpa’ testplan prompts with a box asking if you are 
ready to adjust potentiometers R196 and 197. These two adjustments 
are located on the front panel of the Agilent E6171A module. Select 
OK, to proceed with the adjustments. If you select CANCEL, the 
final manual adjustments to the module are not made.
Figure 6-2. Installing the Calibration Fixture on an Agilent E6171A Measurement Control Module
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com










