Specifications
Calibration and Preventive Maintenance  103Chapter 6
2. Run the ‘autoadj.tpa’ testplan.
3. When prompted by the ‘autoadj.tpa’ testplan, adjust 
potentiometer R196
1
 (front panel of the Agilent E6171B module) to 
match the indicated reading. See Figure 6-3.
4. Select the OK button in the adjustment box. Note: select the 
CANCEL button if you are unable to center the vertical bar.
5. When prompted by the ‘autoadj.tpa’ testplan, adjust 
potentiometer R197
1
 (front panel of the Agilent E6171B module) to 
match the indicated reading. See Figure 6-4.
6. Select the OK button in the adjustment box. Note: select the 
CANCEL button if you are unable to center the vertical bar.
7. System adjustment is complete.
1. Potentiometers R196 and R197 are very sensitive. Typically you should rotate the controls only a slight 
amount. Rotating clockwise moves the bar to the right, rotating counterclockwise moves the bar to the left.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com










