Specifications
System Instrument Configuration  43Chapter 2
Agilent E6171B
Measurement
Control Module
 The Agilent E6171A/B Measurement Control Module (MCM, also known 
as the VI card in the Agilent TS-5400 Action documentation) provides four 
primary functions:
1. It provides a 16x5 matrix used to expand the four columns from the 
32-Pin Matrix Module(s) and UUT Common to access up to 16 
external instruments. 
2. It provides an isolated V/I
1
 programmable voltage and current source 
with internal wiring that allows the external Agilent E1411B DMM to 
measure the sourced voltage or current. The isolated V/I amplifier 
may also be used to amplify signals from the Arbitrary Function 
Generator (ARB) and attenuate signals for digitizing.
3. It routes triggers bi-directionally between the UUT and external 
instruments. 
4. It provides a step-up transformer to amplify the ARB's signal on 
VISrcHi to ± 80 volts-peak to simulate VRS signals.
2
16x5 Matrix The instrument matrix provides 16 external connections which can be 
connected to the four analog buses or to the UUT common ground bus. The 
UUT common ground bus may be an isolated ground return or switched 
through separate relays to V/I ground. It can be connected to system ground 
for diagnostic purposes.
The inputs for isolated instruments, such as the Agilent E1411B Multimeter, 
occupy two instrument channels. Single-ended instruments, like the Agilent 
E1333A Counter, occupy one instrument channel and are referenced to earth 
ground. Rows one through ten are assigned to specific instruments. The 
Agilent E1411B DMM must always be attached to rows one through four 
(DvmHi, DvmLo, DVMISrcHi, DVMISrcLo). If an ARB is used with the 
system to source current or voltage, it must be connected to inputs five and 
six. If a Digitizer (Analog to Digital Converter (ADC)) is used with the 
system, it must be connected to rows seven and eight for port 3 differential 
input, if the attenuator (Attn) is used. The factory default is to use rows nine 
and ten for the Digitizer port 4 differential input. Ports 1 and 2 on the 
digitizer are the single-ended inputs. The counter channel one goes to row 
eleven and channel two to row twelve. Rows thirteen through sixteen can be 
used for additional instrumentation. The Debug Panel ports are connected 
directly to interconnect columns one through four and are referenced to 
system ground. If any of the above instruments are not used in a specific 
system, a recommended factory-default configuration is used for each row 
on instrument MUX connector J1. 
1. V/I is an abbreviation for the concept of sourcing a voltage and measuring the resulting current, or conversely, 
sourcing a current and measuring the resulting voltage. It sometimes generically refers to the Agilent E6171 
Measurement Control Module.