Technical data
8 Service and Maintenance  
Constant Voltage (CV) Verifications
202 E364xA User’s and Service Guide
2 Turn on the power supply. Select the high voltage range 
(20 V/1.545 A)
[1]
, enable the output, and set the display 
to the limit mode. When the display is in the limit mode, 
program the current to the maximum programmable value 
(1.545 A)
[2]
 and the voltage to the full rated value  
(20.0 V)
[1]
.
3 Check that the front panel CV annunciator remains lit. If 
it is not lit, adjust the load down slightly.
4 Note that the waveform on the oscilloscope must not 
exceed the peak- to- peak limit of (5 mV)
[3]
 or (8 mV)
[4]
.
5 Disconnect the oscilloscope and connect the AC rms 
voltmeter in its place according to the 
“Vrms 
measurement techniques”  on page  200 and as shown in 
Figure  8- 4. The rms voltage reading must not exceed the 
rms limit of 0.5 mV
[3]
 or 1 mV
[4]
.
6 Repeat  step  1 through step  5 for the other output.
Load transient response time
This test measures the time for the output voltage to recover 
to within 15 mV of nominal output voltage following a load 
change from full load to half load, or half load to full load.
1 Turn off the power supply and connect the output to be 
tested as shown in 
Figure  8- 2 with an oscilloscope. 
Operate the electronic load in the constant current mode.
2 Turn on the power supply. Select the output to be tested 
and the high voltage range (20 V/1.545 A)
[1]
, enable the 
output, and set the display to the limit mode. When the 
display is in the limit mode, program the current to the 
maximum programmable value (1.545 A)
[2]
 and the 
voltage to the full rated value (20.0 V)
[1]
.
[1] For the E3646A model. For other models, refer to Tabl e  8 -5 .
[2] For the E3646A model. For other models, refer to Tabl e  7 -3 .
[3] For the E3646A/E3648A model.
[4] For the E3647A/E3649A model.










