Technical data
8 Service and Maintenance  
General Calibration or Adjustment Procedure
220 E364xA User’s and Service Guide
15 Save the change and select the high current calibration 
point.
I HI 2.800,0
16 Enter the computed value (DVM reading divided by the 
shunt resistance) by using the knob and resolution 
selection keys.
I HI 2.789,0
17 Save the new current calibration constants for the output1 
and go to the calibration mode for the output2.
VOLT2 CAL
If the calibration fails, an error message will be displayed 
for one second and you will hear a beep, and then return 
k to the current calibration mode again.
18 Repeat  step  3 through step  16 for the voltage, OVP, and 
current calibration.
19 Exit the calibration mode.
CAL MODE
The message above appears if the calibration is successful. 
If the calibration fails, an error message will be displayed 
for one second and you will hear a beep, and then return 
to the current calibration mode again.