User`s guide
Table Of Contents

Language Dictionary 39
OUTP:PROT
There are two output protection commands that do the following:
OUTP:PROT:CLE
Clears any OV (overvoltage), OC (overcurrent, unless set via external voltage control),OT
(overtemperature), or RI (remote inhibit) protection features. After this command, the output is
restored to the state it was in before the protection feature occurred.
OUTP:PROT:DEL
Sets the time between the programming of an output change that produces a CV, CC, or
UNREG condition and the recording of that condition by the Status Operation Condition
register. The delay prevents the momentary changes in power supply status that can occur
during reprogramming from being registered as events by the status subsystem. Since the delay
applies to CC status, it also delays the OCP (overcurrent protection) feature. The OVP
(overvoltage protection) feature is not affected by this delay.
Examples OUTP:PROT:CLE OUTPUT:PROTECTION:CLEAR
OUTPUT:PROTECTION:DELAY 75E-1
OUTP:PROT:DEL MIN OUTPUT:PROT:DELAY MAX
Query Syntax OUTP:PROT:CLE (None)
OUTPut:PROTection:DELay? OUTPut:PROTection:DELay? MIN
OUTPut:PROTection:DELay? MAX
Returned Parameters <NR3> OUTP:PROT:DEL? returns value of programmed delay.
OUTP:PROT:DEL? MIN and OUTP:PROT:DEL? MAX return
the minimum and maximum programmable delays.
Related Commands OUTP:PROT:CLE (None)
OUTP:PROT:DEL *RST *RCL *SAV
OUTP:REL
Do not install or program the Agilent Relay Accessories if the power supply maximum output current
rating (see Table 3-1) exceeds the contact ratings of the relay.
This command is valid only if the power supply is configured for the optional relay connector. Programming ON closes the
relay contacts; programming OFF opens them. The relay is controlled independently of the output state. If the power
supply is supplying power to a load, that power will appear at the relay contacts during switching. If the power supply is not
configured for the relay option, sending either relay command generates an error.
Command Syntax OUTPut:RELay[:STATe] <bool>
Parameters 0 | 1 | OFF | ON
*RST Value 0
Examples OUTP:REL 1 OUTP:REL OFF
Query Syntax OUTPput:RELay?
Returned Parameters 0 |1
Related Commands OUTP[:STAT] *RCL *SAV