12 Verification
Constant Voltage (CV) Tests
Test Setup
Connect your dc voltmeter leads to only +S and -S (see Figure 2-1), because the power supply regulates the voltage
between these points, not between the + and - output terminals.
Test Procedures
Perform the test procedures in Table 2-4. The CV tests are:
Voltage Programming and Readback Accuracy
CV Load Effect
CV Source Effect
CV Noise (PARD)
Transient Recovery Time
Note The tests are independent and may be performed in any order.
a) CV Test Setup with 6.6kW Electronic Load
S
+
S
-
DVM
-
+
-
+
Electronic Load
b) CV Test Setup with 6.6kW Load Resistor
S
+
S
-
DVM
-
+
Note: Start with switch closed, take full load reading,
open switch and take no load reading.
6.6kW Load
Resistor
Figure 2-1. Constant Voltage (CV) Test Setup