Service manual
 Verification 15
 Table 2-4. Constant Voltage (CV) Tests (continued)
 Action
 Normal Result
 Transient Recovery (cont)
 3  Program the Electronic Load as follows:
•  Operating mode to constant current.
•  Input load current to 1/2 the supply's full rated output current.
•  Transient current level to the supply's full rated output current.
•  Transient generator frequency = 100Hz.
•  Transient generator duty cycle = 50%.
 4  Turn on the transient and adjust the oscilloscope to display response
waveform.
 See Fig. 2-2.
 5  Measure both the loading and unloading transients by triggering the
oscilloscope on both the negative and positive slopes of the transient.
Record the voltage level obtained at the 900-µs interval .
 Specified voltage level is reached within
900µs.
 Figure 2-2. Transient Response Waveform
Constant Current (CC) Tests
 Test Setup
 Connect the appropriate current monitoring resistor (see Table 2-l) as shown in Figure 2-3. The accuracy of the resistor
must be as specified in the table.
 Test Procedures
 The test procedures are given in Table 2-5. The tests are independent and may be performed in any order. The CC tests are:
•  Current Programming and Readback Accuracy.
•  CC Load Effect.
•  CC Source Effect.