Technical data
DISPlay Commands SCPI Commands Reference
Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007 35
DISPlay Commands
The following table shows the Agilent 81133A/81134A Pulse 
Generator DISPlay commands.
:DISPlay[:WINDow][:STATe]
Syntax
:DISP[:WIND][:STAT][?]
Form Set & Query
Description This command is used to turn the front panel display on and off. 
Switching off the display improves the programming speed of the 
instrument.
NOTE The display is switched back on if a key on the instrument is 
pressed. The command 
*RST
 switches the display back on. Use 
:SYSTem:PRESet
 to perform a 
*RST
 without switching the display 
back on.
Parameter ON|OFF|1|0
*RST value ON
Example Switch off the front panel display.
:DISP OFF
Command Parameter Description
:DISPlay
 [:WINDow]
 [:STATe][?] ON | OFF | 1 | 
0
Sets/reads front panel display 
state