User`s guide
Setting Up Generic and Advanced Signals Setting Up a Repetitive Burst Signal
52 Agilent 81133A/81134A Pulse Generator User’s Guide, September 2008
Setting Up a Repetitive Burst 
Signal
Task  Set up a signal with 4 repeated bursts. 
Each burst consists of a 32- bit pattern signal with NRZ data 
output format at 100 MHz to check a digital circuit with LVPECL 
logic. 
The data pattern is:
11100000000000000000000000000000
Using the Graphical User Interface
Instrument Settings To set the instrument parameters: 
1 Disable the outputs.
2 In the instrument panel, choose the RBurst mode. 
3 Enter  4 for the number of null data packets.
4 Define that the number of zeros that follows the repeated bursts 
is as large as the burst length.
5 Set the frequency to 100 MHz. 
Data Data Data Data Data Data Data
Data Length = 32 x 10 ns
Burst Length = 4
 x Data Length
Number of Null data packets =
Burst Length