Technical data
Testing Performance 3
16800 Series Portable Logic Analyzers Service Guide 27
Logic Analyzer Test Strategy
Only specified parameters are tested. Specifications are 
listed on page 15. The test conditions defined in this 
procedure ensure that the specified parameter is as good as 
or better than specification. No attempt is made to 
determine performance which is better than specification. 
Not all channels of the logic analyzer will be tested; rather a 
sample of channels is tested. The calibration laboratory may 
choose to elaborate on these tests and test all channels at 
their discretion.
To perform a complete test on a logic acquisition module, 
start at the beginning of the chapter and follow each 
procedure.
Test Interval
Test the performance of the module against specifications at 
two- year intervals.
Test Record Description
For recording the results of each procedure, see 
“Performance Test Record" on page 68.
Test Equipment
Each procedure lists the recommended test equipment. You 
can use equipment that satisfies the specifications given. 
However, the procedures are based on using the 
recommended model or part number.
Instrument Warm-Up
Before testing the performance of the module, warm- up the 
logic analyzer and the test equipment for 30 minutes.