To Test the Minimum Master to Master Clock Time and Minimum Eye Width
The specifications for the logic analyzer define a minimum 
master to master clock time and a minimum data eye width 
at which data can be acquired. This test verifies that the 
logic analyzer meets these specifications. 
Eye finder is used to adjust the sampling position on every 
tested channel. Eye finder must be used to achieve minimum 
data eye width performance.
First, the logic analyzer will be tested in the 250 Mb/s state 
mode. Then it will be tested in the 500 Mb/s state mode. 
In the 250 Mb/s state mode each pod will be tested with its 
respective clock. 
The 500 Mb/s mode has only one clock (Clk1). All tests in 
the 500 Mb/s mode will use clock Clk1.
A sample of four channels on each pod will be tested, one 
pod at a time, in both 250 Mb/s state mode and 500 Mb/s 
state mode. 
The logic analyzer will be configured to acquire data on both 
edges of the clock, so the test frequency is set to half of the 
acquisition speed.