Technical data
Testing Performance 3
16800 Series Portable Logic Analyzers Service Guide 41
Verify and adjust 8133A pulse generator DC offset
1 On the 54845A oscilloscope, select Measure from the menu 
bar at the top of the display.
2 Select  Markers...
3 In the Markers Setup dialog set marker “Ay” to 0.7 V, and 
set marker “By” to 1.3 V.
4 Observe the waveforms on the oscilloscope display. If they 
are not centered within the “Ay” and “By” markers, adjust 
the 8133A pulse generator’s Channel 2 OFFSET until the 
waveforms are centered as close as possible. (The 
resolution of the 8133A OFFSET setting is 10 mV.) 