Technical data
42 16800 Series Portable Logic Analyzers Service Guide
3 Testing Performance
Deskew the oscilloscope
This procedure neutralizes any skew in the oscilloscope’s 
waveform display.
1 On the 54845A oscilloscope, change the Horizontal scale 
to 200 ps/div. You can do this using the large knob in the 
Horizontal setup section of the front panel. 
2 Select  Setup from the menu bar at the top of the display.
3 Select  Channel 1.
4 Select  Probes.
5 Click Skew </> to deskew Channel 1 and Channel 2 signals 
so that both channels cross the 54845A horizontal center 
line at the same time, at both ends of the eye (both 
crossings of the horizontal center line). The horizontal 
center of the graticule line is at 1 volt because the 
vertical offset was set to 1 volt in the oscilloscope setup 
described on page 38.