3 Testing Performance
13 Adjust the sampling positions using eye finder. Be sure to 
expand “My Bus 1” and use the recommended starting 
position noted on page 49. Realign any stray channels if 
necessary. See page 50.
14 Determine pass or fail (1 of 2 tests). See page 52. 
15 Click  OK to close the “Analyzer Setup” dialog.
16 Switch to the Listing window by selecting the Listing tab 
at the bottom of the main logic analyzer window.
17 Click the Run Repetitive icon  .
18 Determine pass or fail (2 of 2 tests). See page 54.
Test the complement of the bits (Pod 2, 250 Mb/s mode)
1 Test the complement of the bits. See page 55.
Test Pods 3 and 4 in 250 Mb/s Mode
1 Perform the normal and complement tests for Pod 3 and 
Pod 4 on the logic analyzer, changing the connection to 
the pod, channel assignments, thresholds, etc. as 
appropriate.
Test using clock Clk3 for Pod 3 and clock Clk4 for Pod 4.