User`s guide
Performance Test Instructions Performance Tests
Agilent 81689A, 81689B and 81649A Tunable Laser Modules User’s Guide, Third Edition 55
13 Note the maximum and minimum values of the calculated Power 
Linearity values for the various settings and record these in the test 
record.
14 Subtract the minimum values from the maximum values of the 
Power Linearity for the various settings. Record these as the Total 
Power Linearity.
Example (Agilent 81689B)
Power Linearity
Maximum Power Linearity at current setting + 0.08 dB
Minimum Power Linearity at current setting - 0.07 dB
Total Power Linearity = (Max Power Linearity - Min Power Linearity)  0.15 dBpp
Power Setting from 
start
Measured Relative 
Power from start
Power reduction 
from start
Power Linearity at 
current setting
Start = REF + 10.00 dBm  0.00 dB  + 0.00 dB =  0.00 dB
+ 9.00 dBm - 1.02 dB + 1.00 dB = - 0.02 dB
+ 8.00 dBm - 1.92 dB + 2.00 dB = + 0.08 dB
+ 7.00 dBm - 3.02 dB + 3.00 dB = - 0.02 dB
+ 6.00 dBm - 3.95 dB + 4.00 dB = + 0.05 dB
+ 5.00 dBm - 5.07 dB + 5.00 dB = - 0.07 dB
+ 4.00 dBm - 5.96 dB + 6.00 dB = + 0.04 dB
+ 3.00 dBm - 7.05 dB + 7.00 dB = - 0.05 dB
+ 2.00 dBm - 8.01 dB + 8.00 dB = - 0.01 dB
+ 1.00 dBm - 8.98 dB + 8.00 dB = + 0.02 dB
0.00 dBm - 9.95 dB + 9.00 dB = - 0.05 dB