User`s guide
Performance Tests Test  Recor d
86 Agilent 81689A, 81689B and 81649A Tunable Laser Modules User’s Guide, Third Edition
Agilent 81649A Performance Test
   Page 6 of 8
Model Agilent 81649A Tunable Laser Report No. ________ Date_______
Power Linearity
Maximum Power Linearity at current setting ______dB
Minimum Power Linearity at current setting ______dB
Total Power Linearity = (Max Power Linearity − Min Power Linearity) ______dBpp
Specification 0.2 dBpp
Measurement Uncertainty  ± 0.05 dB
Power Setting from 
start
Measured Relative 
Power from start
Power reduction 
from start
Power Linearity at 
current setting
Start = REF + 6.00 dBm   0.00 dB  + 0.00 dB =  0.00 dB
+ 5.00 dBm dB + 1.00 dB = dB
+ 4.00 dBm dB + 2.00 dB = dB
+ 3.00 dBm dB + 3.00 dB = dB
+ 2.00 dBm dB + 4.00 dB = dB
+ 1.00 dBm dB + 5.00 dB = dB
− 0.0dBm dB + 6.00dB = dB
− 1.0dBm dB + 7.00dB = dB
− 2.0dBm dB + 8.00dB = dB
− 3.0dBm dB + 9.00dB = dB