User`s guide

Getting Started with Return Loss Making a Return Loss Measurement
22 Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
The Agilent 81000BR Reference Reflector allows you to make a
Reflectance Calibration and a Termination Calibration.
These procedures are decribed in “Calibration using the
Agilent 81000BR Reference Reflector” on page 23.
The Agilent 81610CC Reference Cable also allows you to make a
Reflectance Calibration. Use the measurement patchcord to make
a Termination Calibration.
To measure insertion loss and the front panel delta of the system
(see “Calculating the Front Panel Delta” on page 41), you must
also measure the insertion loss of the Reference Cable and
measurement patchcord.
These procedures are decribed in “Calibration using the Agilent
81610CC Reference Cable” on page 26.