Specifications
Frequency Response Above 3 GHz Performance Test
Above 3 GHz the flatness specification widens out such that higher uncertainties can be 
tolerated, therefore a splitter calibration is not necessary.
Required Test Equipment
Test Equipment Model Number
E4440A 
E4443A, E4445A
E4446A
E4448A
Synthesized Sweeper
83630A/B
83640A/B, 83650A/B
X  
Synthesized Sweeper 83640A/B, 83650A/B   X
Power Meter E4419A/B  X X
Power Sensor 8485A  X  
Power Sensor 8487A    X
Power Splitter 11667B  X  
Power Splitter 11667C   X
3.5 mm Cable 8120-4921 X  
BNC Cable 10503A X X
2.4 mm Cable 8120-6164   X
2.4 mm (f) to 3.5 mm (f) adapter 11901B X  
Type N (m) to 3.5 mm (m) adapter 1250-1743 X  
3.5 mm (f) to 3.5 mm (f) adapter 83059B X  
Type N (m) to 2.4 mm (f) adapter 11903D   X
Power Meter Calibration (Channel A)