Specifications
Measurement Uncertainty Analysis – Instrument Accuracy Test
Measurement Introduction
A-4 Agilent E4416A/E4417A Service Guide
The Measurement Uncertainty is not calculated from different sensor and power 
meter combinations, rather it is taken care of in the Customer Limit specification 
setting. Hence using the same power meter and making measurements a number of 
times will give us the uncertainty of the test station, without the uncertainty 
associated with the meter itself. By taking n samples of the same meter/DUT in effect 
we are averaging out the error associated with the DUT’s measurement capability 
leaving us with a raw test station uncertainty which is made up from TSE and the 
DMM error, for example, the Measured Voltage (MV).
The method that is used to calculate the Measurement Uncertainty of our test station 
is a combination of Type A and Type B analysis in line with the ISO Expression of 
Measurement Uncertainty Guide.
Uncertainty Equation
TSE
The Measurement Uncertainty of the Measurement Accuracy is calculated from 
testing the same DUT at 5dB step power level from –25dBm to +20dBm.
Hence: Measurement Accuracy = f(MV , TSE)
u
2
MA() C
TSE
u TSE()⋅[]
2
C
MV
u MV()⋅[]
2
+=