Specifications
Performance Tests
Power Reference Level Test
Agilent E4418B/E4419B Service Guide 2-17
Equipment
• Test power meter: Agilent 432A.
• Thermistor mount: Agilent 478A option H75 or H76.
• Digital voltmeter (DVM): Agilent 3458A.
Test Setup
Figure 2-4: Power Reference Level Test Setup
Procedure
Use the following procedure to perform the power reference level test: 
Note  The procedure details the key presses required on the 
Agilent E4418B. For the Agilent E4419B the equivalent key presses 
should be performed on both channels.
1. Set the DVM to measure resistance:
Connect the DVM between the V
RF
 connector on the rear panel of the 
Agilent 432A and pin 1 on the thermistor mount end of the sensor cable. 
Verify that no power is applied to the Agilent 432A.
2. Round off the DVM reading to two decimal places and note this value 
as the internal bridge resistance (R) of the Agilent 432A 
(approximately 200 Ω).
•R ____________Ω
3. Connect the equipment as shown in Figure 2-4, ensuring that the DVM 
input leads are isolated from chassis ground.
Note  Ensure that the Agilent 432A and the power meter under test have 
been powered on for at least 30 minutes before moving to step 4.
Power
ref
Power Meter
Thermistor 
mount
Test Power Meter
V
RF 
Vcomp 
Digital Voltmeter
+ input
-input 