Specifications
Performance Tests
Agilent E-Series Power Sensor Interface Test
Agilent E4418B/E4419B Service Guide 2-19
Agilent E-Series Power Sensor Interface Test
The Agilent E-series power sensors have their sensor calibration tables 
stored in EEPROM which enables the frequency and calibration factor 
data to be downloaded automatically by the power meter. The frequency 
and calibration factor data have checksums which are compared with the 
data downloaded by the power meter. For the Agilent E4419B dual 
channel power meter, this test should be performed on both channels.
Procedure
1. When there is no power sensor connected to the Channel A input 
the message “No Sensor” (“No Sensor ChA” on the 
Agilent E4419B) appears on the measurement window. 
2. Connect the Agilent E-series power sensor to the power meter’s 
Channel A input. The message “Reading Sensor ChA” appears. 
When all the data is downloaded this message disappears and a 
reading is displayed on the measurement window. If the download 
fails, a warning message and error flag are displayed on the front 
panel. Also, SCPI error message “-310, System Error” is 
generated.
3. For the Agilent E4419B repeat these steps for Channel B.