Network Hardware User Manual

17
Verification kits
Verification kits are used to verify the performance specifications of an 8510
system. All kits include a precision Z
o
airline, mismatched airline and fixed
attenuators. Measured data and uncertainties traceable to the U.S. National
Institute of Standards and Technology (NIST) are included with each kit.
Compliance with MIL-STD 45662A is available for an extra charge
Choose a verification kit for each connector type required.
Verification Connector Frequency Verification Connector Frequency
kit type range (GHz) kit type range (GHz)
85051B 7 mm 0.045 - 18 R11645A WR-28 26.5 - 40
85053B 3.5 mm 0.045 - 26.5 Q11645A WR-22 33 - 50
85055A Type-N 0.045 - 18 U11645A WR-19 40 - 60
85057B 2.4 mm 0.045 - 50 V11645A WR-15 50 - 75
W11645A WR-10 75 - 110
Test port cables and adapters
Test port cables and adapter sets are available for various connector types.
Special test port adapter sets convert the rugged ports of the network analyzer
test set to the desired connector interface. Each kit contains two adapters, one
male and one female. Both the cables and the test port adapters have one spe-
cial female connector which is designed to connect directly to the 3.5 mm test
port (2.4 mm for 8517B). This side of the cable or adapter can only be connected
to the test set port and cannot be mated to a standard 3.5 mm (or 2.4 mm)
male connector. Choose one of the configurations shown.
Configuration A. This cable arrangement is for applications where the device
under test is connected directly to the test set port. This setup offers the best
mechanical rigidity for device connection. To adapt the test set port (port 1) to
the device under test, choose the appropriate special adapter set. In addition
to converting the test port to the desired interface, these adapters also func-
tion as “test port savers” which protect the test set from damage and wear due
to heavy use.
For Agilent 8514B/8515A/85110A test sets (3.5 mm rugged test port connectors)
Cables/adapter Connector type
(on device side of cable/adapter)
For 3.5 mm devices 85131C semi-rigid cable or 3.5 mm (f)
85131E flexible cable 3.5 mm (f )
85130D adapter set 3.5 mm (m and f)
For 7 mm devices 85132C semi-rigid cable or 7 mm
85132E flexible cable 7 mm
85130B adapter set 7 mm
For Type-N devices Use 7 mm cables and the 7 mm-to-Type-N
adapters included in the 85054B/D
Type-N calibration kit
For Agilent 8517B test sets (2.4-mm rugged test port connectors)
Cables/adapter Connector type
(on device side of cable/adapter)
For 2.4 mm devices 85133C semi-rigid cable or 2.4 mm (f)
85133E flexible cable 2.4 mm (f )
85130G adapter set 2.4 mm (m and f)
For 3.5 mm devices 85134C semi-rigid cable or 3.5 mm (f)
85134E flexible cable 3.5 mm (f )
85130F adapter set 3.5 mm (m and f)
For 7 mm devices 85135C semi-rigid cable or 7 mm
85135E flexible cable 7 mm
85130E adapter set 7 mm
Network analyzer
test set
Test port
adapter
Device
under
test
Single cable
Configuration A