Specifications
14
Innovative 
Applications
Fast and accurate 
noise figure  
measurements
(Option 028, 029) 
continued
PNA-X’s unique source-corrected noise figure solution
• Uses modified cold-source method, eliminating need for noise source when measuring DUT
• Corrects for imperfect system source match by using vector correction to remove mismatch 
errors plus an ECal module used as an impedance tuner to remove noise-parameter- 
induced errors
• Maintains high measurement accuracy in fixtured, on-wafer, or automated-test environments
• Accurately measures differential devices using vector 
deembedding of baluns or hybrids
DUT 
Measure differential devices by 
deembedding baluns or hybrids.
At each test frequency, four or more noise measurements 
are made with known, non-50-ohm source impedances. 
From these measurements, 50-ohm noise figure is 
accurately calculated.
DUT  
Test port 1 
R1 
Test port 2 
R2 
A 
B 
To receivers 
LO 
Source 2 
Output 1 
Source 2 
Output 2 
Pulse generators 
REAR PANEL 
1 
2 
3 
4 
+28 V 
Noise receivers 
10 MHz to 
3 GHz 
3 to
26.5 GHz 
Source 1 
OUT 1  OUT 2 
Pulse 
modulator 
Source 2 
OUT 1  OUT 2 
Pulse 
modulator 
Noise source used 
for calibration only 
J9 J10 
J11 
J8 
J7 
J2  J1 
Impedance 
tuner for 
noise figure
measurements
+ 
– 
Block diagram of a two-port N5242A PNA-X with Options 
200, 219, 224, and Noise Figure Option 029. A standard 
ECal module is used as an impedance tuner to help 
remove the effects of imperfect system source match. 
N5244/45/47A models include a built-in impedance tuner.