Specifications
2. IF Gain
Uncertainty
Specification
Related Performance
Test
Description
0.0 dBm to -55.9 dBm . . . . . . . . . . . . . . . . . . . . . . . . . . f0.6 
dB
-56.0 dBm to -129.9 dBm
. . . . . . . . . . . . . . . . . . . . . . 
hl.0
 dB
IF Gain Uncertainty Test
A signal source of known amplitude is connected to the spectrum
analyzer and the analyzer is adjusted for a reference level. The
amplitude of the signal peak is measured in 0.1 
dB
 steps from -0.1 
dB
to -1.9 
dB,
 in 2 
dB
 steps from -1.9 
dB
 to -9.9 
dB,
 and in 10 steps
from -10 dB to 
-
 120 
dB.
5-12 Operation Verification