User`s guide

lo.
Press Define Gate. Use the GATE DELAY and GATE LENGTH keys to position the gate.
Once gate delay or gate length are activated, use the knob and data entry keys can be used
to position the two vertical gate markers. Select a time interval within the last half of the
pulse is selected.
11. Turn the gate on by pressing GATE ON OFF so that ON is underlined. This activates the
frequency domain window, which is the lower window. The spectrum selected with
the current gate position can now be viewed in the frequency domain window. Press
Main Menu.
12. The resolution bandwidth, video bandwidth, and sweep time are not optimized, so the
frequency display may not look correct. There may be signal dropouts or poor frequency
resolution. This can be corrected by entering the pulse parameters and turning on the
coupling.
The gate utility can optimize the setting of resolution bandwidth if the user enters the
value of the pulse width and turns on the coupling. The video bandwidth will be optimized
if the gate length is entered and coupled. The sweep time is optimized when the pulse
repetition interval is entered and coupled.
Press Define
Coupling .
Then press Pulse
Param
to enter the pulse parameters. (This
activates the time domain window and turns off the time gate.) If pulse parameters have
previously been entered, the values will be displayed.
12.
Use the ENTER REF EDGE, ENTER WIDTH, and ENTER PRI
softkeys
to enter the pulse
parameters. These parameter entry tools allow pulse parameters to be entered using a
marker or through the keypad. Press Previous Menu to return to the coupling menu.
14.
Press CPL RBW ON OFF (ON) to turn on the resolution bandwidth coupling. Press
CPL VBW ON OFF (ON), and CPL SWP ON OFF (ON) to turn on the video bandwidth and
sweep time coupling.
15.
Press Main Menu and look at the signal in the time domain window.
16.
Press UPDATE
TIMEFREIJ
so that FREQ is underlined or press
mEXT),
to activate the
frequency window instead of the time domain window. (If the gate was not on when
the user left the frequency window, it may be necessary to press Define Gate and
GATE ON OFF (ON) to turn the gate on again.)
Note
another front panel key), press the key twice to return to the last gate
utility menu used.
Making Measurements
4.21