User`s guide

Completing
‘Ihble
Entry and Activating Limit-Line Testing
Pressing EDIT DONE blanks the limit-line table from the screen and accesses the menu with
LMT TEST ON OFF and LMT DISP Y N AUTO softkeys.
Pressing LMT TEST UN
OFF
turns the limit-line testing on and off.
Pressing LMT DISP Y N AUTO (Y) turns the limit lines display on. Pressing
LMT DISP Y N AUTO (N) turns the limit lines display off. Pressing LMT DISP Y N AUTO
(AUTO) sets the limit line display to match the limit line test function. With AUTO underlined
the limit lines are only displayed when limit line testing is turned on.
Saving or Recalling Limit-Line
‘lhbles
Pressing
[DISPLAY)
then Limit Lines accesses SAVE LIMIT and RECALL LIMIT. These
softkey
functions provide an easy way to save or recall current limit-line tables. SAVE LIMIT
saves the current limit-line tables in the current mass storage device (either spectrum analyzer
memory or a memory card). To determine the current mass storage device, press SAVE LIMIT .
If MAX REG
#
appears on the spectrum analyzer display, the current mass storage device is
analyzer memory. If PREFIX= is displayed, the memory card is the mass storage device. Enter
a register number, then press
@Ki%i?j
to save the current limit-line table in the current mass
storage location.
RECALL LIMIT recalls limit-line tables from the current mass storage device (either spectrum
analyzer memory or a memory card). To determine the current mass storage device, press
RECALL LIMIT . If MAX REG # appears on the spectrum analyzer display, the current mass
storage device is analyzer memory. If PREFIX= is displayed, the memory card is the mass
storage device. To recall a limit line, enter the register number that the limit-line table was
saved under, then press
~j%K@.
When recalling a limit line from the memory card, it is
necessary that the current prefix matches the prefix that the limit line was stored with. Use
Change Prefix to change the current prefix.
5-28 Using Analyzer Features