User`s guide

Amplitude-Correction Functions
This section describes the amplitude-correction functions in the order they are usually used.
Editing or Viewing the Amplitude-Correction
‘lhbles
Pressing
COAL),
More 1 of 4 ,
More 2 of 4, More 3 of 4, Amp Cor accesses the
softkey
menus for creating an amplitude-correction table.
Note
(PRESET) turns amplitude-correction factors off (if it is on), but does not clear an
existing amplitude-correction table. Use PURGE AMP COR to clear an existing
amplitude-correction table.
Press Edit Amp Cor to edit an existing amplitude-correction table or, if no
amplitude-
correction table currently exists, to create an amplitude-correction table.
Pressing PURGE AMP COR two times consecutively clears an existing amplitude-correction table.
Selecting the Amplitude-Correction Point
Pressing SELECT POINT specifies the amplitude-correction point to be entered or edited.
Amplitude-correction data is constructed from left to right and is created by entering
frequency and amplitude values into an amplitude-correction table. The frequency and
amplitude values specify a coordinate point from which amplitude-corrections are interpolated
(see Figure
5-13).
Up to 79 points can be specified for the amplitude-corrections table.
16:24:30
1.7 MAR 1992
&
REF
.B
dl3rn
ATTEN
iE
dB
PEAK
LOG
18
dB/
STOP
400.0 MHz
:
I
STOP 400.0 MHz
RES BW
3.0 MHz UBW 1 MHz
SWP
20.6
msec R
Figure 5-13. Amplitude-Correction Points
Item
Description of Items in Figure 5-13
1
Frequencies below first point use first amplitude level.
2
First segment interpolated with the 10
dB
amplitude level.
3
Frequency and amplitude coordinate that starts the second segment.
4
Third segment interpolated with the
-
10
dB
amplitude level.
5
Frequency and amplitude coordinate that starts the fourth segment.
5-38
Using Analyzer Features