User`s guide

ALL DLP
-
CARD
ALC TEST
HP 8590L or HP 8591 E with Option 010 or 011 only.
breaks the leveling loop of the automatic leveling control in the tracking
generator. This is a service diagnostic function and is for service use only.
Refer to the service guide for more information.
Front-Panel Key Access:
ICAL)
Requires Option 003 for an HP
859OL,
HP
8592L,
or HP 8594L.
saves all the downloadable programs and key definitions that are in spectrum
analyzer memory onto the memory card. If the downloadable program was
stored using a prefix, the file name for the downloadable program consists of
d(prefix)-(register number). If no prefix was specified, the data is stored with
the file name d-(register number).
Front-Panel Key Access:
ISAVE_)
Amp Cor
AMP COR
accessed by
ICAL).
Amp Cor accesses the menus for entering and editing the
current amplitude-correction factors.
Front-Panel Key Access:
a
when accessed by
m,
AMP COR stores the current amplitude-correction
factors table in spectrum analyzer memory or on the memory card.
When accessed by
(j-j,
AMP COR recalls the amplitude-correction
factors table from either spectrum analyzer memory or the memory card.
Amplitude-correction factors are saved with an “a” before the memory-card
file name. Screen titles are not recalled with the data. Refer to “To Save
a Limit-Line Table or Amplitude Correction Factors” in Chapter 5 for more
information.
Amplitude-correction-factor memory-card tiles can be
catalogued
using
CATALOG AMP COR.
Front-Panel Key Access:
[RECALL)
or
B
AMP COR
ON OFF
(AMPLITUDE)
turns the current amplitude-correction factors on or off.
Front-Panel Key Access: (CAL)
activates the reference level function and accesses the amplitude menu. The
softkeys
accessed when you press
Cm)
change reference level, input
attenuation, vertical scale, mixer level, amplitude units, input impedance,
and amplitude offset. For the HP 85933, HP 85943, HP 85953, or HP
8596E,
pressing
cm)
accesses the preselector peaking and preselector default
functions also.
Amptd
Units
accesses the
softkeys
that change the amplitude units. The amplitude units can
be changedby pressing
dBm,
dBmV,
dBuV,
Volts, or Watts.
Front-Panel Key Access:
[AMPLITUDE)
7-8
Key Descriptions