User`s guide

Band
Lock
HP
8592L,
HP
8593E,
HP
85954
and HP 8596E only.
accesses the harmonic band menu and the band lock function. Selecting
a harmonic band causes the spectrum analyzer to lock onto the specified
harmonic band and automatically select the settings shown in Table 7-2.
‘able
7-2.
Center Frequency and Span Settings for Harmonic Bands
Softkey
Center
Span
Frequency
O-2.9
Gz
BAND 0
1.45
GHz
2.9
GHz
2.75-6.5 BAND 1
4.638
GHz
3.6
GHz
6.0-12.8 BAND 2 *
9.4
GHz
6.8
GHz
12.4-19.4 BAND 3
t
15.9
GHz
7
GHz
19.1-22 BAND 4
t
20.55
GHz
2.9
GHz
* Not available with the HP 85953.
t
Not available with the HP 85953 or HP 85963.
Description
Low-pass filtered, first harmonic mixing.
Preselected, first harmonic mixing.
Preselected, second harmonic mixing.
Preselected, third harmonic mixing.
Preselected, fourth harmonic mixing.
BND LOCK
ON OFF
locks the spectrum analyzer onto a selected frequency band (local
oscillator harmonic number). When only one frequency band is being swept
the corresponding
softkey
will be underlined, even if band lock is off.
Note
When using the spectrum analyzer in a band lock mode, the span is limited
to 3.6
GHz
in band 0 and 1, and to 7
GHz
in bands 2 through 4. To select the
maximum span in a given band, use the start frequency, stop frequency, or
span function.
Front-Panel Key Access:
(FREQUENCY)
BAUD
RATE
Q&ion
043
(AS-232 and parallel interface) only.
sets the data transmission speed. (Also see the description for the IcoPv] key.)
The baud rate is set to 1200 by pressing DEFAULT
CONFIG
.
Front-Panel Key Access:
(E%iZj
BINARY
SPAN
HP
85901,
and HP 85913 only.
displays the output of the span DAC that is located on the A7 Analog Interface
assembly. This is a service diagnostic function and is for service use only.
Front-Panel Key Access: (CAL)
BLANK A
stores the amplitude data for trace A and removes it from the screen. The
trace A register will not be updated as the spectrum analyzer sweeps.
Front-Panel Key Access:
@EXE]
BLANK B
stores the amplitude data for trace B and removes it from the screen. The
trace B register will not be updated as the spectrum analyzer sweeps.
Front-Panel Key Access:
[TRACE)
BLANK C
stores the amplitude data for trace C and removes it from the screen. The
trace C register will not be updated as the spectrum analyzer sweeps.
Front-Panel Key Access: @EiiE]
7-l
2
Key Descriptions