User`s guide

PRT MENU
Option
041
or 043 only.
ON OFF
allows the
softkey
labels to be printed along with the spectrum analyzer
display. This function operates when the (copy) key is used in a print
configuration. The PRT MENU function is set to ON when DEFAULT
CONFIG
is
pressed.
Front-Panel Key Access:
C-j
Pulse
Param
PURGE
AMP
COR
PURGE
LIMITS
Option 105 only. Option 101 is recommended.
accesses the gate utility menus for entering the pulse parameters: reference
edge, pulse width, and pulse repetition interval. If pulse parameters have
previously been entered, pressing Pulse
Param
causes a list of the current
values to be displayed.
Front-Panel Key Access:
(SWEEP’
clears the current amplitude-correction factors table. Pressing PURGE AMP COR
displays the message: If you are sure, press key againtopurge data.
Pressing PURGE AMP COR a second time clears the amplitude-correction data.
Press SAVE AMP COR to save amplitude-correction factors, and then press
PURGE AMP COR to clear the current amplitude-correction factors table.
Front-Panel Key Access: (CAL)
clears the current limit-line table from spectrum analyzer memory. Pressing
PURGE LIMITS displays the message: If you are sure, press key again
to purge data. Press PURGE LIMITS againifyou wish to clear the current
limit-line table. Press SAVE LIMIT to save the current limit-line table, and
then press PURGE LIMITS to clear the current limit-line table.
Front-Panel Key Access:
(mj
PWRGRAPH
ON OFF
turns the channel power graph ON or OFF. With the PWRGRAPH ON, the
channel powergraph is calculated and displayed and the numeric results are
not displayed. The value of the channel power is displayed at the selected
marker frequency. This graph function is used after doing a channel power
measurement with the CHANNEL POWER softkey.
Front-Panel Key Access:
(MEAS/USERJ
7-70 Key Descriptions