User`s guide

(SAVE_)
SAVE
AMP COR
accesses
softkey
menus that allow you to store state data, trace data, limit-line
tables, and amplitude-correction factors on a memory card or in spectrum
analyzer memory. The SAVE function also allows you to save state data, trace
data, limit-line tables, amplitude-correction factors, and program data on the
memory card. In addition, pressing
m
accesses the
softkey
menus used to
catalog the saved data in spectrum analyzer memory or on the memory card.
To save to, or catalog from, spectrum analyzer memory press INTERNAL CARD
so that INTERNAL is underlined. To save to, or catalog from, the memory
card, press INTERNAL CARD so that CARD is underlined. The HP
859OL,
HP
8592L,
and HP 8594L must have Option 003 for memory card operation.
Saving state data saves the spectrum analyzer settings, but not the trace data.
Saving trace data saves both the trace data and the state data. Display images
and programs (also called downloadable programs or
DLPs),
can only be saved
to or recalled from the memory card.
States and traces are saved in spectrum analyzer memory even if
the instrument is turned off or
(-1
is pressed. Eight spectrum
analyzer-memory state registers and many trace registers are available for the
user. The Catalog Internal
softkey
is used to access the catalog functions.
It also accesses the DLP editor utility.
saves the current amplitude-correction factors table to the current
mass-storage device (spectrum analyzer memory or memory card). To verify
the current mass storage device, press SAVE AMP COR . If MAX REG # appears
on the spectrum analyzer display, the current mass storage device is spectrum
analyzer memory. If PREFIX= is displayed, the memory card is the mass storage
device. Press
(SAVE_)
or
cm),
then INTERNAL CARD to change the current
mass storage device. Press SAVE AMP COR , enter a register number, then press
@YZj
to save the current amplitude-correction factors table in spectrum
analyzer memory or on the memory card. When saved on the memory card,
amplitude-correction factors tables are stored with “a-“, the prefix, and the
register number entered. When saved in spectrum analyzer memory, the
register number is saved in a trace register. Trace register values are restricted
to a range between 0 and the number x indicated by MAX
REG#
= x.
Front-Panel Key Access:
ICALl
SAVE
EDIT
Fbr Option 041 or 043 only. Refer to the HP 8590 E-Series and L-Series
Spectrum Analyzer and HP 8591 C Cable TV Analyzer Programmer’s Guide for
mare information.
passes the text from the DLP editor memory through the parser to execute as
spectrum analyzer commands. Pressing SAVE EDIT is similar to outputting
the text to the spectrum analyzer from an external controller. If the text
(commands) is a valid user-defined function, it passes through the parser and
into the spectrum analyzer user memory. It will replace an existing user
defined function of the same name.
The DLP editor memory buffer remains intact when the spectrum analyzer is
preset or powered off so the text being edited will not be lost.
Front-Panel Key Access:
@i?ZiJ
or
LSAVE)
Key Descriptions 7-75