Technical data

10
86100A/B/C
Firmware Release Notes
A.06.00 Release Notes, December 15, 2005 (Differences from Release
A.04.20)
New Feature List
Added Enhanced Impedance and S-parameters (#202)
Support for S-parameter results for single-ended (S11, S21, S12, and S22), differential
(SDD11, SDD21, SDD22, SDD12), and common modes (SCC11, SCC21, SCC22, SCC12).
S-parameter display graphs feature markers, graph formatting, Autoscale, and print
support
S-parameter is calculated between two time periods which allows customers to using time
gating before converting to S-parameter
Corrected impedance profile for single-ended and differential signals
Allows export of S-parameter measurements into either .s1p or .s2p Touchstone data
format for import into circuit simulators
Support for 86100C Mainframe
New TDR User Interface for measurement setup and display
Improved TDR Calibration that supports time base changes, time delays, and zooming into
data
Guided TDR Calibration Procedure to improve usability
Support of external TDR edge accelerators in TDR Mode
Extensive support for math functions like derivatives, constant multiplications, add,
subtract, and scaling
Maximum/Minimum/Average Measurements in TDR Mode
Export of time data in X & Y format
Display of ER Cal factor
Changed from hardware graphics to software graphics
General Firmware Defects Fixed in A.06.00
Increased maximum allowable time base delay in TDR mode to 100 ns.
86100B + 54754A previously did not invert white trace color on print. The waveform color
is set to gray during printing to prevent a white waveform trace on a white background.
Repaired 86106B not successfully completing Autoscale with filter on
Resolved ?g:DISK:SIM 'filename.jpg' ?gcausing instrument to hang when file specified is
read-only. Added error checking before attempting to save file.
Phase is correct now for reference clock frequencies between 25 GHz and 39 GHz on the
40 GHz input of an 86107A Option 40 module.
Added a remote command to raise the jitter shade :DISPlay:JITTer:SHADe
Allow TDR mode to accept maximum time base delay up to 100ns
Extinction Ratio calibration factor is always loaded into dialog boxes
After 86100C power cycle, TDR mode is recalled properly
Known Issues
PLTS customers should not upgrade to A.06.00 software at this time since the executable
will fail to run properly. PLTS customers should use DCA firmware version A.04.11.
86119A (Optical Sampling Oscilloscope) is designed to work on software versions A.03.06
and A.03.05 only.
86100A/B Mainframes are not supported on DCA Firmware Revision A.06.00