Technical data
10
86100A/B/C
Firmware Release Notes
A.06.00 Release Notes, December 15, 2005 (Differences from Release
A.04.20)
New Feature List
• Added Enhanced Impedance and S-parameters (#202)
• Support for S-parameter results for single-ended (S11, S21, S12, and S22), differential
(SDD11, SDD21, SDD22, SDD12), and common modes (SCC11, SCC21, SCC22, SCC12).
• S-parameter display graphs feature markers, graph formatting, Autoscale, and print
support
• S-parameter is calculated between two time periods which allows customers to using time
gating before converting to S-parameter
• Corrected impedance profile for single-ended and differential signals
• Allows export of S-parameter measurements into either .s1p or .s2p Touchstone data
format for import into circuit simulators
• Support for 86100C Mainframe
• New TDR User Interface for measurement setup and display
• Improved TDR Calibration that supports time base changes, time delays, and zooming into
data
• Guided TDR Calibration Procedure to improve usability
• Support of external TDR edge accelerators in TDR Mode
• Extensive support for math functions like derivatives, constant multiplications, add,
subtract, and scaling
• Maximum/Minimum/Average Measurements in TDR Mode
• Export of time data in X & Y format
• Display of ER Cal factor
• Changed from hardware graphics to software graphics
General Firmware Defects Fixed in A.06.00
• Increased maximum allowable time base delay in TDR mode to 100 ns.
• 86100B + 54754A previously did not invert white trace color on print. The waveform color
is set to gray during printing to prevent a white waveform trace on a white background.
• Repaired 86106B not successfully completing Autoscale with filter on
• Resolved ?g:DISK:SIM 'filename.jpg' ?gcausing instrument to hang when file specified is
read-only. Added error checking before attempting to save file.
• Phase is correct now for reference clock frequencies between 25 GHz and 39 GHz on the
40 GHz input of an 86107A Option 40 module.
• Added a remote command to raise the jitter shade :DISPlay:JITTer:SHADe
• Allow TDR mode to accept maximum time base delay up to 100ns
• Extinction Ratio calibration factor is always loaded into dialog boxes
• After 86100C power cycle, TDR mode is recalled properly
Known Issues
• PLTS customers should not upgrade to A.06.00 software at this time since the executable
will fail to run properly. PLTS customers should use DCA firmware version A.04.11.
• 86119A (Optical Sampling Oscilloscope) is designed to work on software versions A.03.06
and A.03.05 only.
• 86100A/B Mainframes are not supported on DCA Firmware Revision A.06.00