Technical data

21
86100A/B/C
Firmware Release Notes
86100A-Only Firmware Release Notes
A.03.01 Release Notes, February 15, 2002 (Differences from Release
A.02.20)
New Feature List
Ability to acquire multiple channels at once in Eye Mode
Ability to make measurement and mask analysis on multiple eyes at the same time
Measurement results tab has a new look: clear meas button replaced by ?gx?h on meas
results tab and tab up/down button replaced by ?g?Þ?h on meas results tab
Parametric mask creation
Ohms/Division setting added for TDR mode
Ability to annotate and delete individual measurements
Add support for the following new modules: 86111U, 86111A, 86115B, 86113A, 86102A,
86103B, 86116A, 86107A, 83494A Options 103 and 107
Defects Fixed:
When filter is turned on, autoscale switches filter when completed for no reason
Trigger setting changes after autoscale
System crash when non-TDR module is autoscaled in TDR mode
Stop Menu item behaves as Stop/Single
Source parameter is not being accepted with 'MEASure' subsystem commands
Sending DISP:PERS INF in eye mode sets it to CGR
Screen capture when in color grade saves files in grayscale
Saving .jpg on full diskette results in 0 record length file
Remote uploaded waveform memories are not stored on disk
Remote command returns mainframe calibrated when it is not; added :CAL:REC
CHAn<N> command
Quickmeasure attempts to measure waveform memories in eye mode
Querying waveform values without waveform displayed hung the GUI
Probe calibration hangs up box after remote command. We no longer initiate optical cal if
chan is electrical.
Need more digits in read-out of X Markers at low time/div setting
MTEST:START fails after running MTEST:LOAD on bad mask name, but displays no
warning
MTEST:SAVE .msk saves as '.msk.pcm ? Measurement Results Tab drops in height when