Technical data

8
86100A/B/C
Firmware Release Notes
A.07.00 Release Notes, March 15, 2007 (Differences from Release
A.06.01)
New Features List
The following features are available only with Option 300, Advanced Amplitude Analysis/
RIN/Q-factor (which requires Option 200, Enhanced Jitter Analysis Software and 86100C
Option 001, Enhanced Trigger): Amplitude Analysis, Separation and display of amplitude
interference components (similar to jitter mode), RIN/RINoma, Q factor
Added Optical Modulation Amplitude measurement to oscilloscope mode.
Support for 83496B clock recovery module with the capability to lock on data streams that
employ a Spread Spectrum Clock (SSC). Enables phase noise analysis on clocks.
Support for Option 83496A-UAB, upgrade to 83496B capability.
Support for new 86116C high speed optical and electrical module. The 86116C offers
reference receiver performance for the OC768 standard.
Upper limit on LBW increased to 24M/n , which allows wider bandwidth for signals with .5
transition density.
Added queries DISK:TFILe? and DISK:BFILe? To transfer text and binary files over GPIB
New setting can automatically turn on Quick Measure when entering Eye/Mask mode or
Oscilloscope mode.
General Defects Fixed
The queries time:pos?, time:range?, time:scale?, wav:xdis?, wav:xran? Now properly
return zoomed parameters.
Waveform display for TDR common mode response now update properly after zoom.
ER calibration delta temp now uses the module temperature.
RZ One and Zero level measurements now make measurement at the peak of the pulse
instead of the center.
DISPlay:DATA? And DISK:SIMage with GRATicule parameter will not include the mask
test results tab in the screen image.
Saving pattern waveform now uses *OPC when finished.
Fixed Differential TDR calibration update problem when changing timebase settings.
Repaired an issue that could occur when trying to load a Calibration file, that does not
exist, over GPIB.
Load jitter data base is no longer available outside jitter mode.
Horizontal Auto Skew functionality improved.
Zoom in DDJ vs. Bit plot is now properly limited to 2 bits.
Repaired an issue that could occur when using LFE or Matlab with no modules present.
Mask test always includes hits from final acquisition when stop is pressed during mask test.
Known Issues
86119A (Optical Sampling Oscilloscope) is designed to work on software versions A.03.06
and A.03.05 only.