User`s guide

1-54 Chapter1
System Verification and Performance Tests
Performance Tests
Power Meter Calibration, Port 1 from 300 kHz to 3 GHz
1. Zero and calibrate the power meter.
2. Connect the equipment as shown in Figure 1-19.
Figure 1-19 Setup for Power Meter Calibration on Port 1
3. Press Preset, Start, 300, k/m.
4. Only for analyzers with Option 006: Press Stop, 3, G/n.
5. Press Local, SYSTEM CONTROLLER.
6. Press SET ADDRESSES and POWER MTR until the analyzer shows the
correct power meter model.
7. Press ADDRESS: P MTR/GPIB. The default power meter GPIB address
is 13. Make sure it is the same as your power meter GPIB address.
Otherwise, use the analyzer front panel keypad to enter the correct
GPIB address for your power meter.