User`s guide

Chapter 1 1-55
System Verification and Performance Tests
Performance Tests
8. Press Sweep Setup, NUMBER of POINTS, 51, x1.
9. Press Power, PWR RANGE MAN to turn the auto power range off.
NOTE The analyzer displays the PRm annotation, indicating that the analyzer
power range is set to MANUAL.
10.Press PORT POWER to uncouple the port output power.
11.Press Cal, PWRMTR CAL.
12.Press LOSS/SENSR LISTS, CAL FACTOR SENSOR A. Refer to the back of
the power sensor to locate the different calibration factor values along
with their corresponding frequencies.
NOTE The analyzer’s calibration factor sensor table can hold a maximum of 55
calibration factor data points.
The following softkeys are included in the sensor calibration factor
entries menu:
SEGMENT Press to select a point where you can use the front
panel knob or entry keys to enter a value.
EDIT Press to edit or change a previously entered value.
DELETE Press to delete a point from the sensor calibration
factor table.
ADD Select this key to add a point into the sensor
calibration factor table.
CLEAR LIST Select this key to erase the entire sensor calibration
factor table.
DONE Select this key when done entering points to the
sensor calibration factor table.