User`s guide

Chapter 1 1-61
System Verification and Performance Tests
Performance Tests
Port 2 Input Frequency Response from 300 kHz to 3 GHz,
Forward Direction
1. Disconnect the RF cable from Port 1.
2. Connect the RF cable to Port 2.
Figure 1-24 Port 2 Input Frequency Response, 300 kHz to 3 GHz
3. Press Meas, SELECT PORTS [2-3].
4. Press Marker Fctn, MARKER MODE MENU, STATS ON.
5. Read the peak-to-peak (p-p) value.
NOTE The deviation between the minimum dB value and the maximum dB
value over the frequency range is the peak-to-peak measurement for this
test.
6. Write the peak-to-peak value on the performance test record.
7. Press Marker Fctn, MARKER MODE MENU, STATS OFF.
8. Press Meas, SELECT PORTS [1-3].
9. If your analyzer does not have Option 006, this completes the
procedure. If you have Option 006 in your analyzer, continue with the
following sections.