User`s guide

Chapter 1 1-63
System Verification and Performance Tests
Performance Tests
8. Press RETURN, TAKE CAL SWEEP to start the power meter calibration.
Port 1 Input Frequency Response from 3 GHz to 6 GHz
1. When the analyzer finishes the calibration sweep, connect the
equipment as shown in Figure 1-26.
Figure 1-26 Setup for Port 1 Input Frequency Response
2. Press Meas, INPUT PORTS, A.
3. Press Marker, MARKER 1, Marker Search, SEARCH:MIN to put
marker 1 at the minimum magnitude location of the trace.
4. Press Marker, MARKER 2, Marker Search, SEARCH:MAX to put marker 2
at the maximum magnitude location of the trace.
5. Write the marker 1 or marker 2 reading, whichever has the largest
absolute magnitude, on the performance test record.