User`s guide
Chapter 1 1-65
System Veriļ¬cation and Performance Tests
Performance Tests
Power Meter Calibration, Port 1 from 3 GHz to 6 GHz
1. Connect the equipment as shown in Figure 1-28.
Figure 1-28 Setup for Power Meter Calibration on Port 1
2. Press Meas, INPUT PORTS, TESTPORT 1.
3. Press Cal, PWRMTR, ONE SWEEP, TAKE CAL SWEEP to start the power
meter calibration for Port 1.