User`s guide
1-66 Chapter1
System Veriļ¬cation and Performance Tests
Performance Tests
Port 3 Input Frequency Response from 3 GHz to 6 GHz
1. When the analyzer displays the message POWER METER
CALIBRATION SWEEP DONE, connect the equipment as shown in
Figure 1-29.
Figure 1-29 Port 3 Input Frequency Response Test Setup
2. Press Meas, INPUT PORTS, B.
3. Press Marker, MARKER 1, Marker Search, SEARCH:MIN to put
marker 1 at the minimum magnitude location of the trace.
4. Press Marker, MARKER 2, Marker Search, SEARCH:MAX to put marker 2
at the maximum magnitude location of the trace.
5. Write the marker 1 or marker 2 reading, whichever has the largest
absolute magnitude, on the performance test record.