User`s guide
Chapter 1 1-67
System Veriļ¬cation and Performance Tests
Performance Tests
Port 2 Input Frequency Response from 3 GHz to 6 GHz,
Reverse Direction
1. Disconnect the RF cable from Port 3.
2. Connect the RF cable to Port 2 as shown in Figure 1-30.
Figure 1-30 Port 2 Input Frequency Response, Reverse Direction
3. Press Meas, SELECT PORTS [1-2].
4. Press Marker Fctn, MARKER MODE MENU, STATS ON.
5. Read the peak-to-peak (p-p) value.
NOTE The deviation between the minimum dB value and the maximum dB
value over the frequency range is the peak-to-peak measurement for this
test.
6. Write the peak-to-peak value on the performance test record.
7. Press Marker Fctn, MARKER MODE MENU, STATS OFF.
8. Press Meas, SELECT PORTS [1-3].