User`s guide

Chapter 1 1-69
System Verification and Performance Tests
Performance Tests
8. Test Port Crosstalk
Specifications
Required Equipment
Analyzer warmup time: 1 hour
Frequency Range
Crosstalk
a
300 kHz to 3 GHz < 100 dB
3 GHz to 6 GHz
b
< 87 dB
a. Measurement conditions: 25 ± 5 °C; normalize to a through;
measured with two shorts (or shielded open) 10 Hz IF BW;
averaging factor 8; alternate model; source power at +10 dBm
b. Only for analyzers with Option 006
Description Model Number or Part Number
Calibration Kit, 50, Type-N 85032B
Adapter, Type-N (m) to Type-N (m) part number 1250-1475
Cable, 50, Type-N, 24-inch part number 8120-4781