User`s guide
Chapter 1 1-83
System Verification and Performance Tests
Performance Tests
Source Match (Forward) Calibration Coefficient
1. Press System, SERVICE MENU, TESTS, 33, x1, EXECUTE TEST.
2. When the analyzer finishes the test, press Scale Ref, AUTO SCALE,
Marker
.
3. Using the front panel knob, locate the maximum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.
4. Write the maximum value on the performance test record.
5. Repeat the previous two steps for the other frequency range(s) listed
on the performance test record.
Reflection Tracking (Forward) Calibration Coefficient
1. Press System, SERVICE MENU, TESTS, 34, x1, EXECUTE TEST.
2. When the analyzer finishes the test, press Scale Ref, AUTO SCALE,
Marker
.
3. Using the front panel knob, locate the maximum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.
4. Write the maximum value on the performance test record.
5. Repeat the previous two steps for the other frequency range(s) listed
on the performance test record.
Load Match (Reverse) Calibration Coefficient
1. Press System, SERVICE MENU, TESTS, 42, x1, EXECUTE TEST.
2. When the analyzer finishes the test, press Scale Ref, AUTO SCALE,
Marker
.
3. Using the front panel knob, locate the maximum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.
4. Write the maximum value on the performance test record.
5. Repeat the previous two steps for the other frequency range(s) listed
on the performance test record.