User`s guide

1-84 Chapter1
System Verification and Performance Tests
Performance Tests
Transmission Tracking (Reverse) Calibration Coefficient
1. Press System, SERVICE MENU, TESTS, 43, x1, EXECUTE TEST.
2. When the analyzer finishes the test, press Scale Ref, AUTO SCALE,
Marker
.
3. Using the front panel knob, locate the maximum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.
4. Write the maximum value on the performance test record.
5. Repeat the previous two steps for the other frequency range(s) listed
on the performance test record.
Port 1 to Port 3 Second Full 2-Port Calibration
1. Connect the equipment as shown in Figure 1-40.
Figure 1-40 Second Full 2-Port Calibration Test Setup (Port 1 to Port 3)
2. Press Preset, Start, 300, k/m.
3. Press Cal, CAL KIT, SELECT CAL KIT, CAL KIT:N 50, RETURN, RETURN,
CALIBRATE MENU, FULL 2-PORT
.
4. Press ISOLATION, OMIT ISOLATION.
5. Connect the open (supplied in the calibration kit) to reference test
port 1.
6. Press REFLECTION, FORWARD:OPEN, OPEN (M), DONE OPENS.