User`s guide

Chapter 1 1-91
System Verification and Performance Tests
Performance Tests
Transmission Tracking (Reverse) Calibration Coefficient
1. Press System, SERVICE MENU, TESTS, 43, x1, EXECUTE TEST.
2. When the analyzer finishes the test, press Scale Ref, AUTO SCALE,
Marker
.
3. Press MARKER 2. Using the front panel knob, locate the maximum
value of the data trace for the 300 kHz to 1.3 GHz frequency range.
4. Press MENU MODE, REF=1. Write the value on the performance
test record.
5. Repeat the previous two steps for the other frequency range(s) listed
on the performance test record.
6. Press MARKER, ALL OFF.
Port 2 to Port 3 Second Full 2-Port Calibration
1. Connect the equipment as shown in Figure 1-44.
Figure 1-44 Second Full 2-Port Calibration Test Setup (Port 2 to Port 3)
2. Press Preset, Start, 300, k/m.
3. Press Meas, SELECT PORTS [2–3].