Chapter 1 1-93
System Verification and Performance Tests
Performance Tests
19.Connect the equipment as shown in Figure 1-45.
Figure 1-45 Transmission Calibration Test Setup (Port 2 to Port 3)
20. Press TRANSMISSION, DO BOTH FWD + REV.
21. Press DONE 2-PORT CAL.
Load Match (Forward) Calibration Coefficient
1. Press System, SERVICE MENU, TESTS, 36, x1, EXECUTE TEST.
2. When the test is done, press Scale Ref, AUTO SCALE, Marker,
MARKER 1
.
3. Using the front panel knob, locate the maximum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.
4. Write the maximum value on the performance test record.
5. Repeat the previous two steps for the other frequency range(s) listed
on the performance test record.
Directivity (Reverse) Calibration Coefficient
1. Press System, SERVICE MENU, TESTS, 32, x1, EXECUTE TEST.
2. When the test is done, press Scale Ref, AUTO SCALE, Marker,
MARKER 1
.