User`s guide
1-94 Chapter1
System Verification and Performance Tests
Performance Tests
3. Using the front panel knob, locate the maximum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.
4. Write the maximum value on the performance test record.
5. Repeat the previous two steps for the other frequency range(s) listed
on the performance test record.
Source Match (Reverse) Calibration Coefficient
1. Press System, SERVICE MENU, TESTS, 39, x1. At the prompt, press
EXECUTE TEST.
2. When the test is done, press Scale Ref, AUTO SCALE, Marker,
MARKER 1
.
3. Using the front panel knob, locate the maximum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.
4. Write the maximum value on the performance test record.
5. Repeat the previous two steps for the other frequency range(s) listed
on the performance test record.
Reflection Tracking (Reverse) Calibration Coefficient
1. Press System, SERVICE MENU, TESTS, 40, x1, EXECUTE TEST.
2. When the test is done, press Scale Ref, AUTO SCALE, Marker,
MARKER 1
.
3. Using the front panel knob, locate the maximum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.
4. Write the maximum value on the performance test record.
5. Repeat the previous two steps for the other frequency range(s) listed
on the performance test record.
Transmission Tracking (Forward) Calibration Coefficient
1. Press System, SERVICE MENU, TESTS, 37, x1, EXECUTE TEST.
2. When the analyzer completes the test, press Scale Ref, AUTO SCALE,
Marker, MARKER 1
.
3. Using the front panel knob, locate the minimum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.