User`s guide

Chapter 1 1-95
System Verification and Performance Tests
Performance Tests
4. Press MARKER 2.
5. Using the front panel knob, locate the maximum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.
6. Press MENU MODE, REF=1. Write the value on the performance
test record.
7. Repeat steps 3 through 5 for the other frequency range(s) listed on
the performance test record.
8. Press Marker, All OFF.
Port 1 to Port 2 Full 2-Port Calibration
1. Connect the equipment as shown in Figure 1-46.
Figure 1-46 First Full 2-Port Calibration Test Setup (Port 1 and Port 2)
2. Press Preset, Start, 300, k/m.
3. Press Meas, SELECT PORTS [1–2].
4. Press Cal, CAL KIT, SELECT CAL KIT, CAL KIT:N 50, RETURN, RETURN,
CALIBRATE MENU, FULL 2-PORT.
5. Press ISOLATION, OMIT ISOLATION.
6. Connect the open (supplied in the calibration kit) to Port 1.
7. Press REFLECTION, FORWARD:OPEN, OPEN (F), DONE OPENS.