User`s guide

1-98 Chapter1
System Verification and Performance Tests
Performance Tests
Load Match (Reverse) Calibration Coefficient
1. Press System, SERVICE MENU, TESTS, 42, x1, EXECUTE TEST.
2. When the analyzer finishes the test, press Scale Ref, AUTO SCALE,
Marker
.
3. Using the front panel knob, locate the maximum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.
4. Write the maximum value on the performance test record.
5. Repeat the previous two steps for the other frequency range(s) listed
on the performance test record.
Transmission Tracking (Reverse) Calibration Coefficient
1. Press System, SERVICE MENU, TESTS, 43, x1, EXECUTE TEST.
2. When the analyzer finishes the test, press Scale Ref, AUTO SCALE,
Marker
. Using the front panel knob, locate the maximum value of the
data trace for the 300 kHz to 1.3 GHz frequency range.
3. Press MARKER 2. Using the front panel knob, locate the maximum
value of the data trace for the 300 kHz to 1.3 GHz frequency range.
4. Press MENU MODE, REF=1. Write the value on the performance
test record.
5. Repeat the previous two steps for the other frequency range(s) listed
on the performance test record.
6. Press Marker, All OFF.